FEATURES
Large 128x64 dots, backlit LCD screen with brightness control displays a large array of data also in graphical format
Large sample compartment to accomodate 5-100mm path length cuvette with optional holders
Auto setting of wavelength-to calibrate the system through arrow keys and avoid operational errors.
The system can save the test results, up to 200 groups of data and 200 standard curves in the RAM memory
Data can also be restored after a sudden power failure
Optical System
Single Beam
Light Source
Tungsten lamp
Spectral Bandwidth
4.0nm
Wavelenth Range
325-1000nm
Wavelength Accuracy
±2nm
Wavelength Repeatability
≤1nm
Photometric Range
-0.3-3.0A, 0-200%T
Photometric Accuracy
±0.003A@ 0.5A, ±0.006A@ 1.0A, ±0.5%T @100%T
Photometric Repeatability
≤0.0015A@ 0.5A, ≤0.003A@ 1.0A, 0.25%T @100%T
Stability
≤0.002A/h @ 500nm
(after preheat 2 hours)
Stray Light
≤0.2%T@ 360nm
Sample Compartment
4-position, 10mm
pathlength cuvette
Display
LCD
Output
USB Port& Parallel Port (Printer)
Power
110/220VAC, 50/60Hz, 80W
Dimensions[LxWxH]
490x360x210nm
Weight
10Kg